Fixed VOC Detector
... extended warranty with long sensor life
Model PIDScan 800
PIDScan 800 is a revolutionary state-of-the-art fixed VOC detectors based on PID technology for continuous measurement of VOCs. It is the world's first fixed type, ex-approved VOC detector with automatic self-cleaning feature. Literally, there is no cleaning of the lamp required. Both zeroing and span calibration can be carried out automatically. The transmitter has resounding performance even in high temperature and humidity conditions; as a result of its unique “Earth-ring guard” technology, and built-in temperature and humidity sensors for real-time compensation.
Features:
Automatic self-cleaning
to ensure reliable, improved life and operation performance of the PID lamp
Automatic
zeroing
for fast repeatable response and
accuracy for long duration
Unique ceramic-glass PID sensor
has proven to yield stable performance and long-sensor life beyond expectations
Built-in temperature and humidity sensors
to ensure accurate and real-time compensation measurements under all weather conditions
-
4-20 mA analog and MODBUS RS485 output ; HART as optional.
-
PID sensor highly resistance to humidity interference
-
Explosion proof EEx d IIC T6 for hazardous application
-
With built-in diaphragm pump to draw in sample gas
-
Magnetic interface for non-intrusive access to menu
-
Large numerical and graphical dot-matrix display with backlight sensor
Specifications:
PID sensor 10.6 eV
Range / Resolution / 0 - 2000 ppm / 1 ppm (high range version)
(optional range) 0.0 - 100.0 ppm / 0.1 ppm (standard version) / (0.0 - 1000 ppm)
0.00 - 10.00 ppm / 0.01 ppm ( ultra sensitive version ) / ( 0.00 - 20 ppm)
* with Temperature and Humidity compensation on board
Self-cleaning Built-in self-cleaning
Relay contacts SPST dry contacts (x3)
Ring LEDs Alarm Built-in ring alarm LEDs; Yellow and Red
Response Time T90 < 10 sec (improved)
EX Approval ATEX II 2G Ex d IIC T6 Gb
UL Class I Div I Group B, C and D
IP Rating IP66 EMI / RFI
Compliance to EMI / RFI Directive 89 / 336 / EEC
Power 9 - 36 Vdc, max. 125 mA at 24 Vdc
Cable 3-wire shielded; max. 1 kOhm at 24 Vdc
Output / (option) 4 - 20 mA , HART (option)
Modbus RS 485
SPST relay contacts (x2), Fault relay (x1); max. 30 V, 2 A.
Sampling method Built-in diaphragm pump (~500 ml / min)
Display Graphic LCD 128 x 64 pixels, with backlit
User interface Non-intrusive using magnetic key
Operating Cond. Temperature: -20 degC to +60 degC (-4 degF to 140 degF) Humidity: 0 - 95%RH (non-condensing)
Physical Data Weight: 3.5 Kg
Dimension: L 124 x D 125 x H 253 (mm)
OTHER PRODUCTS
Head Office (Singapore)
Site Navigation
GDS INSTRUMENTS PTE LTD
33 Ubi Avenue 3, Unit 05-28,
Vertex Building Tower B
Singapore 408868.
Email: sales@gds-instruments.com
Tel: +(65) 6509 4758 (Hunting)
Fax: +(65) 6509 4227
Support Centre
Manuals
Product Catalogue
Marketing Bulletins
Service and Support
Customer Feedback
Legal
Privacy Policy
Terms and Conditions
Disclaimer